Command Reference
R&S
®
ZNC
588
User Manual 1173.9557.02 ─ 13
[SENSe<Ch>:]FREQuency:CONVersion:GAIN:LMCorrection
<Boolean>
Enables or disables the load match correction for frequency conversion transmission S-
parameters (conversion gain factors).
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
*RST:
ON
Example:
See
[SENSe<Ch>:]FREQuency[:CW]
<FixedFreq>
[SENSe<Ch>:]FREQuency:FIXed
<FixedFreq>
Defines the fixed (Continuous Wave, CW) frequency for all sweep types operating at fixed
frequency (power sweep, time sweep, CW mode sweep). The two command forms
[SENSe<Ch>:]FREQuency[:CW]
and
[SENSe<Ch>:]FREQuency:FIXed
are equiv-
alent.
The frequency range is listed in
.
Note:
[SENSe<Ch>:]FREQuency[:CW]|:FIXed
is equivalent to
SOURce<Ch>:FREQuency<PhyPt>[:CW]|:FIXed
. Source and receiver frequency
are always equal; the four commands overwrite each other.
Suffix:
<Ch>
.
Channel number
Parameters:
<FixedFreq>
Fixed stimulus and analyzer frequency
Increment: 0.1 kHz
*RST:
1 GHz
Default unit: Hz
Example:
FUNC "XPOW:POW:A1"
Activate a power sweep and select the wave quantity a
1
as mea-
sured parameter for channel and trace no. 1.
FREQ:CW 100MHz
Set the CW frequency to 100 MHz.
Manual operation:
See
[SENSe<Ch>:]FREQuency:SBANd
<Sideband>
Defines whether the analyzer measures with a local oscillator frequency LO below or
above the RF input frequency.
SCPI Command Reference