Command Reference
R&S
®
ZNC
582
User Manual 1173.9557.02 ─ 13
[SENSe<Ch>:]CORRection:PCAL
<ActivatePowerCals>
Activates or deactivates all power calibrations.
Suffix:
<Ch>
.
Calibrated channel number
Setting parameters:
<ActivatePowerCals> NONE | ALL
NONE
–
deactivate all power calibrations.
ALL
–
activate all power calibrations.
*RST:
n/a
Example:
*RST; CORR:POW:ACQ AWAVe,1,PORT,1
Perform a receiver power calibration of the wave a
1
using port 1
as a source port, assuming that the transmitted source power a
1
is correct. No external cabling is needed.
CORR:POW:AWAV?
Check whether the calibration is applied (the response is
1
).
CORR:PCAL NONE; POW:AWAV?
Disable all power calibrations and check again whether the cali-
bration is applied (the response is
0
).
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:POWer<PhyPt>:AWAVe[:STATe]
<Boolean>
Enables or disables the receiver power calibration for channel
<Ch>
and for the reference
waves a
n
. The setting command is disabled unless the reference waves have been power
AWAVe,...
). The query always returns a result.
This command extends the functionality of manual control, where the reference receiver
calibration is included in the source power calibration of each port (and always ON).
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the receiver power calibration for
the reference waves a
n
.
*RST:
OFF
Example:
See
SCPI Command Reference