Command Reference
R&S
®
ZNC
528
User Manual 1173.9557.02 ─ 13
<Format>
COMPlex | LINPhase | LOGPhase
COMPlex - complex values (real and imaginary part)
LINPhase - linear magnitude and phase.
LOGPhase - dB-magnitude and phase.
If the second optional parameter is omitted, the command stores
complex data.
<DecSeparator>
POINt | COMMa
POINt - Decimal separator: point.
COMMa - Decimal separator: comma.
If the third optional parameter is omitted, points are used.
<FieldSeparator>
SEMicolon | COMMa | TABulator | SPACe
SEMicolon - Field separator: semicolon
COMMa - field separator: comma.
TABulator - field separator: tabulator.
SPACe - field separator: space.
If the third optional parameter is omitted, semicolons are used.
Example:
*RST; :CONF:TRAC:NAME?
Reset the instrument, creating a default channel no 1 and a default
trace Trc1.
CALC:PAR:DEF:SGR 1,2
Create four traces to measure the two-port S-parameters S
11
,
S
12
, S
21
, S
22
. The traces are not displayed.
MMEM:STOR:TRAC:CHAN 1, 'C:\Users\Public
\Documents\Rohde-Schwarz\Vna\Traces\Chn1.csv'
Store all trace data of channel 1 to a trace file.
MMEM:STOR:TRAC:CHAN 1, 'C:\Users\Public
\Documents\Rohde-Schwarz\Vna\Traces\Chn1.s2p'
Store the four S-parameter traces to a two-port Touchstone file.
The Touchstone file will not contain the default trace Trc1.
Usage:
Setting only
Manual operation:
See
MMEMory:STORe:TRACe:PORTs
<Channel>, <TouchstoneFile>, <Format>, <Port>,
<Port>...
Generates a Touchstone file for the specified ports. The Touchstone file (.s<n>p where
<n> is the number of ports) contains a full set of <n>2 single-ended S-parameters for the
selected ports. Traces are created using the
command.
The command fails unless the conditions for Touchstone file export are met; see
ditions for Touchstone file export"
Setting parameters:
<Channel>
Channel number in the active recall set.
SCPI Command Reference