Command Reference
R&S
®
ZNC
522
User Manual 1173.9557.02 ─ 13
MMEMory:STORe:CKIT
<CalKitName>, <CalKitFile>
Stores the data of a calibration kit to a specified file. The calibration kit is identified by its
name.
Setting parameters:
<CalKitName>
Name of a user-defined calibration kit available on the analyzer.
Tip:
It is not possible to modify or store predefined or ideal kits.
<CalKitFile>
String parameter to specify the name and directory of the cal kit
file to be created. The file is a network analyzer-specific cal kit file
with the extension
*.calkit
. If no path is specified the analyzer
uses the current directory, to be queried with
.
Example:
MMEM:LOAD:CKIT 'C:\Users\Public\Documents
\Rohde-Schwarz\Vna\Calibration\Kits\New Kit
1.calkit'
Load the previously created cal kit file
New Kit 1.calkit
from
the default cal kit directory.
... :MMEM:STOR:CKIT 'New Kit 1', 'C:
\Users\Public\Documents
\Rohde-Schwarz\Vna\Calibration\Kits\New Kit
1.calkit'
Store the data for the user-defined cal kit "New Kit 1" and overwrite
the cal kit file
New Kit 1.calkit
.
Usage:
Setting only
Manual operation:
See
MMEMory:STORe:CKIT:WLABel
<CalKitName>, <Label>, <CalKitFile>
Stores the data of a calibration kit to a specified file. The calibration kit is identified by its
name and label.
Parameters:
<CalKitName>
Name of a user-defined calibration kit available on the analyzer.
Tip:
It is not possible to modify or store predefined or ideal kits.
<Label>
Label of the calibration kit, usually its serial number.
<CalKitFile>
String parameter to specify the name and directory of the cal kit
file to be created. The file is a NWA-specific cal kit file with the
extension *.calkit. If no path is specified the analyzer uses the cur-
rent directory, to be queried with
Example:
See
Usage:
Setting only
SCPI Command Reference