GUI Reference
R&S
®
ZNC
267
User Manual 1173.9557.02 ─ 13
The diagram in the center of the dialog visualizes the settings and results below.
Access:
CHANNEL > CAL > Pwr Cal Settings > Cal Power...
Port Overview
The dialog shows all source ports of the network analyzer. Each port is displayed with
the current "Power Result" at the input of the DUT (in dBm) and offset (i.e. the "Cal Power
Offet" in dB).
Remote command:
Port Power Offset
Defines a port-specific offset to the channel base power. The actual output power at the
port is equal to the channel base power P
b
("CHANNEL > POWER BW AVG > Power
> ...") plus the "Port Power Offset". It is equal to the "Port Power Offset" (converted into
dBm) if "0 dBm" is selected instead of P
b
.
For power sweep and the selection P
b
, the actual port power varies across the sweep;
for other configurations the port power is constant.
Remote command:
Channel Settings