
Command Reference
R&S
®
ZNB/ZNBT
858
User Manual 1173.9163.02 ─ 55
CALCulate<Chn>:LIMit:DCIRcle:DISPlay[:STATe]
<Boolean>
Sets/queries the visibility of the display circle for the active trace of channel <Chn>.
The display circle is defined using
CALCulate<Chn>:LIMit:DCIRcle:DATA
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
If set to ON, the line of the display circle is shown whenever the
related trace is displayed in complex format.
Manual operation:
See
CALCulate<Chn>:LIMit:DELete:ALL
Deletes all limit line segments.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
*RST; :CALC:LIM:CONT 1 GHZ, 1.5 GHZ
Define an upper limit line segment in the stimulus range
between 1 GHz and 1.5 GHz, using default response values.
CALC:LIM:DATA 1,1500000000, 2000000000,2,3
Define an upper limit line segment in the stimulus range
between 1.5 GHz and 2 GHz, assigning response values of +2
dB and +3 dB.
CALC:LIM:DEL:ALL
Delete both created limit line segments.
Usage:
Event
Manual operation:
See
"Add / Insert / Delete / Delete All"
CALCulate<Chn>:LIMit:DISPlay[:STATe]
<Boolean>
Displays or hides the entire limit line (including all segments) associated to the active
trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON | OFF - Limit line on or off.
*RST:
OFF
Example:
*RST; :CALC:LIM:CONT 1 GHZ, 2 GHZ
Define an upper limit line segment in the stimulus range
between 1 GHz and 2 GHz, using default response values.
CALC:LIM:DISP ON
Show the limit line segment in the active diagram.
Manual operation:
See
SCPI Command Reference