GUI Reference
R&S
®
ZNB/ZNBT
674
User Manual 1173.9163.02 ─ 55
Embedded LO
Select "Embedded LO" if in the related mixer measurement the LO cannot be provided
by the VNA or a connected external generator. Setting this flag makes the LO drive
power range read-only.
Use Fixed Freq
Uncheck "Use Fixed Freq" if you want the LO signal to vary in frequency, which will
enable "LO Freq Min" and "LO Freq Max". Otherwise only "LO Fixed Freq" is enabled.
6.15.4
Distance to Fault Application
With option K3, the R&S
ZNB/ZNBT can locate faults and discontinuities on cables and
transmission lines. The measured S
ii
-parameter trace is mathematically converted to
the time domain, and represented in the distance domain. The x-axis of the DtF trace
(derived from S
ii
) shows the distance from the calibrated reference plane; faults appear
as peaks above a certain level.
If option K3 is not installed on the R&S
ZNB/ZNBT, the "Distance to Fault" application
is hidden.
To achieve the best distance resolution, set the sweep span as large as possible and
use the
to adjust the number of sweep points.
Background information
Refer to
Chapter 5.7.3, "Distance to Fault Measurements"
6.15.4.1
Controls on the Distance to Fault Tab
The controls on the "Distance to Fault" tab allow you to enable and configure a stan-
dard Distance to Fault (DtF) measurement.
Applic Softtool