GUI Reference
R&S
®
ZNB/ZNBT
333
User Manual 1173.9163.02 ─ 55
"Real"
Dyn_Shift(z) = z – Re(zr)
"Imag"
Dyn_Shift(z) = z – Im(zr)
"Complex"
Dyn_Shift(z) = z – zr
Remote command:
n.a.
6.5.10
Infinite Averaging Tab
The controls on the "Infinite Averaging " tab allow you configure infinite averaging for
the active trace.
Infinite averaging means the average at a sweep point is taken over all trace values at
this sweep point since the previous restart of the averaging cycle.
Infinite Average/Restart
"Infinite Average" enables or disables "Infinite Averaging " (default: disabled), "Restart"
starts a new averaging cycle.
Remote command:
CALCulate<Chn>:IAVerage[:STATe]
Mode
Selects the quantities to be averaged.
"Magnitude -
Phase"
Perform averaging of magnitude and phase (default) of the complex
trace value.
Real - Imag
Perform averaging of the real and imaginary part of the complex trace
value (i.e. over the complex value itself).
Remote command:
6.5.11
Trace Data Tab
Stores one or several data or memory traces to a file or loads a memory trace from a
file.
Trace Config Softtool