Product Overview
R&S
®
TSVP
20
Getting Started 1152.3943.62 ─ 07
●
"Folded matrix" concept used :
– Odd channels (P1, P3, P5) can be connected directly to odd analogue
busses (A1,B1,C1,D1)
– Even channels can be connected directly to even analogue bus lines
(A2,B2,C2,D2)
– Odd and even channels can be connected using so called bus sense
relays
●
GND pins available to separate DUT GND lines for ICT ("GND decoupling")
●
3 Instrument lines from front connector X10 to analogue bus and Rear I/O
●
Cross talk attenuation:
≤
-80 dB at 10 kHz,
≤
-40 dB at 1 MHz
Each R&S TS-PMB module can be used in R&S CompactTSVP and R&S
PowerTSVP.
3.3.11
R&S TS-PSAM
The R&S TS-PSAM Analog Stimulus and Measurement Module is used to per-
form analog measuring tasks, in-circuit measurements, and the R&S Com-
pactTSVP self-test. The module's floating DC supply voltage is provided via the
associated R&S TS-PDC rear I/O module. This is inserted in the appropriate rear
I/O slot in the back of the device (same slot number as R&S TS-PSAM).
It offers the following key features:
●
Floating measurement inputs using analogue bus matrix and front connector
●
Digital multimeter:
Measurement ranges:
– 10 mV - 125 V VDC, 20 mV - 90V VAC (RMS)
– 1
μ
A - 1A IDC, 100
μ
A - 1A, IAC (RMS)
– 1
Ω
- 10 M
Ω
(2- and 4-wire cablings possible)
16-Bit A/D Converter, max. 200 ksamples/s
●
Source:
Floating DC source : - 5 V to + 5 V, max. 100mA (DCS)
●
Discharge:
On-module discharge unit up to 125V and 400mA (Discharge)
Test Modules