Remote Control Commands
R&S
®
SMJ100A
746
Operating Manual 1403.7458.32 ─ 14
[:SOURce]:LFOutput:VOLTage
<Voltage>
Sets the voltage of the LF output signal.
Parameters:
<Voltage>
float
Range:
see the data sheet: Internal modulation generator >
Output voltage range
Increment: see the data sheet: resolution of output voltage set-
ting
*RST:
1
Example:
LFO:VOLT 2 V
sets the voltage of the LF output to 2 V.
Manual operation:
See
7.16.11 SOURce:LIST Subsystem
This subsystem contains the commands for the List mode of the instrument.
The following settings are required to operate the instrument in List mode:
1. Create a list.
If a list which does not exist is selected with the
:LIST:SEL
command, an empty
list with the name of the selected list is created.
SOURce1:LIST:SEL "New_list"
2. Fill the list with values.
All list components must be of the same length. This does not apply to components
of length 1. This is interpreted as if the component has the same length as the
other components and as if all values are the same as the first value.
SOURce1:LIST:FREQ 100 MHz, 110 MHz, 120 MHz...
SOURce1:LIST:POW 2dBm, -1dBm, 0dBm...
3. Select a list.
If a new empty file has been created with the
:LIST:SEL
command, this file is
selected, otherwise an existing list must be selected before the List mode is activa-
ted.
SOURce1:LIST:SEL "Old_list"
4. Set the dwell time.
The dwell time determines the duration of the individual list steps.
SOURce1:LIST:DWELl 3ms
5. Set the List mode.
The List mode determines the way in which the list is processed. In the example
the list is processed once only or repeatedly depending on the trigger setting.
SOURce1:LIST:MODE AUTO
6. Determine the trigger.
SOURce Subsystem
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...