Remote Control Commands
R&S
®
SMJ100A
631
Operating Manual 1403.7458.32 ─ 14
Parameters:
<Source>
ZERO | ONE | PRBS | PATTern | DLISt | SERial | PARallel
ZERO
An internally generated 0 data sequence is used.
ONE
An internally generated 1 data sequence is used.
PRBS
The pseudo-random sequence generator is used as the data
source.The length of the random sequence is defined with the
aid of command
SOURce:BB:DM:PRBS
.
PATTern
Internally generated data with a fixed pattern is used. The data
pattern is dafined using command
SOURce:BB:DM:PATTern
.
DLISt
Data from the selected data list is used.
SERial
External data from the serial interface is used.
PARallel
External data from the serial interface is used.
*RST:
PRBS
Example:
BB:DM:SOUR DLIS
the internal data generator is used.
BB:DM:DLIS:SEL 'test'
the data list
test.dm_iqd
is used.
Manual operation:
See
[:SOURce<hw>]:BB:DM:SRATe
<SRate>
Sets the symbol rate. The value range is dependent on the selected modulation type.
On changing to an FSK modulation type, excessively high values are automatically
limited to the maximum value that can be set for FSK (see data sheet). The symbol
rate can be entered in Hz/kHz/MHz or Symb/s / kSymb/s and MSymb/s.
When a standard is selected (
DM:STANdard
), the symbol rate is automatically set to
the appropriate default value.
The value range is dependent on the selected modulation type (:
BB:DM:FORM
). When
a standard is selected (
:BB:DM:STAN
), the symbol rate is set to the default value.
Parameters:
<SRate>
float
Range:
400 to depends on hardware
Increment: 0.001
*RST:
270833.333
Default unit: Hz
Example:
BB:DM:SRAT 10 MHz
sets a symbol rate of 10 MHz.
SOURce Subsystem
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...