Remote Control Commands
R&S
®
SMJ100A
607
Operating Manual 1403.7458.32 ─ 14
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:PHASe[:STARt]
<Start>
The command sets the start phase for the individual carriers in the defined carrier
range. If the command
:BB:ARB:MCAR:EDIT:CARR:PHAS:STEP
is used to define a
step width, the phase entered here applies only to the starting carrier. The phases of
the remaining carriers are stepped up or down by the phase value specified in
the
:BB:ARB:MCAR:EDIT:CARR:PHAS:STEP
command.
The phase settings are only valid if optimization of the crest factor is disabled
(
:SOURce:BB:ARB:MCARrier:CFACtor:MODE OFF
).
Parameters:
<Start>
float
Range:
0 to 359.99
Increment: 0.01
*RST:
0
Default unit: DEG
Example:
BB:ARB:MCAR:EDIT:CARR:PHAS 90 DEG
sets a start phase of 90° for the carriers in the carrier range.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer:STEP
<Step>
The command sets the step width by which the starting power of the carriers in the
defined carrier range will be incremented.
Parameters:
<Step>
float
Range:
-80 to 80
Increment: 0.01
*RST:
0
Default unit: dB
Example:
BB:ARB:MCAR:EDIT:CARR:POW -80dB
sets a power of -80 dB for the carriers in the carrier range.
BB:ARB:MCAR:EDIT:CARR:POW:STEP 1 dB
the power is incremented by 1dB for each carrier, i.e. the first
carrier has -80dB, the second -79dB, etc.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer[:STARt]
<Start>
The command sets the power for the individual carriers in the defined carrier range. If
the command
:BB:ARB:MCAR:EDIT:CARR:POW:STEP
is used to define a step width,
the power entered here applies only to the starting carrier. The power of the remaining
carriers is stepped up or down by the power specified in
the
:BB:ARB:MCAR:EDIT:CARR:POW:STEP
command.
SOURce Subsystem
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...