Remote Control Commands
R&S
®
SMJ100A
501
Operating Manual 1403.7458.32 ─ 14
Setting parameters:
Mode
UNCHANGED
The segments may have different clock rates.
During the multi segment calculation, the clock rates of all indi-
vidual original waveforms were taken over unchanged in the
segments of the multi segment output waveform (no software
resampling is done).
If the segments have different clock rates, there are some
restrictions on signal output, i.e. switching per external trigger
and seamless switching between segments is not possible.
If all segments have already an identical clock rate, this mode is
usually set to
HIGHEST
.
HIGHEST
All segments have an indentical clock rate, which is the highest
clock rate of all original waveforms.
If an individual original waveform has a lower clock rate, an
upsampling to the highest clock rate will performed on calcula-
tion of the multi segment output waveform.
USER
All segments have an indentical clock rate, which is given by the
user.
Note: Only upsampling is allowed, no downsampling!
If an individual original waveform has a lower clock rate, an
upsampling to the user clock rate will performed on calculation
of the multi segment output waveform.
Example:
{MWV_SEGMENT_CLOCK_MODE: UNCHANGED}
{MWV_SEGMENT_CLOCK_MODE: HIGHEST}
{MWV_SEGMENT_CLOCK_MODE: USER}
Usage:
Setting only
{MWV_SEGMENT_CLOCK: ClockSeg0, ClockSeg1, …, ClockSegN-1}
(mandatory for multi segment waveforms)
The tag contains a list of clock frequencies for every segment in the multi segment
waveform in ASCII floating point format.
Example:
{MWV_SEGMENT_CLOCK: 100e6,80e6}
2 segments: clock of segment 0 is 100 MHz, clock of segment 1
is 80 MHz.
Note: If the segments have different clock frequencies, there are
some restrictions on signal output, i.e. seamless switching
between segments is only possible, if all segments have the
same clock frequency. Software resampling (upsampling) can
be used to bring all segments to the same clock.
Usage:
Setting only
Waveform and List Format
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...