![R&S SMJ100A Скачать руководство пользователя страница 461](http://html1.mh-extra.com/html/rands/smj100a/smj100a_operating-manual_3086611461.webp)
Remote Control Basics
R&S
®
SMJ100A
445
Operating Manual 1403.7458.32 ─ 14
●
A wired "OR"-connection is used if several instruments are connected in parallel,
since the slowest instrument determines the speed.
Any connected IEC bus cable must be terminated by an instrument or controller.
6.1.4.1
GPIB Interface Messages
Interface messages are transmitted to the instrument on the data lines, with the atten-
tion line (ATN) being active (LOW). They are used for communication between the
controller and the instrument and can only be sent by a computer which has the func-
tion of a GPIB bus controller. GPIB interface messages can be further subdivided into:
●
Universal commands
: act on all instruments connected to the GPIB bus without
previous addressing
●
Addressed commands
: only act on instruments previously addressed as listeners
Universal Commands
Universal commands are encoded in the range 10 through 1F hex. They affect all
instruments connected to the bus and do not require addressing.
Command
Effect on the instrument
DCL (Device Clear)
Aborts the processing of the commands just received and sets the com-
mand processing software to a defined initial state. Does not change the
instrument settings.
IFC (Interface Clear) *)
Resets the interfaces to the default setting.
LLO (Local Lockout)
The LOC/IEC ADDR key is disabled.
SPE (Serial Poll Enable)
Ready for serial poll.
SPD (Serial Poll Disable)
End of serial poll.
PPU (Parallel Poll Unconfig-
ure)
End of the parallel-poll state.
*) IFC is not a real universal command, it is sent via a separate line; however, it also affects all instruments
connected to the bus and does not require addressing
Addressed Commands
Addressed commands are encoded in the range 00 through 0F hex. They only affect
instruments addressed as listeners.
Command
Effect on the instrument
GET (Group Execute Trigger)
Triggers a previously active instrument function (e.g. a sweep).
The effect of the command is the same as with that of a pulse at
the external trigger signal input.
GTL (Go to Local)
Transition to the "local" state (manual control).
GTR (Go to Remote)
Transition to the "remote" state (remote control).
Remote Control Interfaces and Protocols
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...