Instrument Function
R&S
®
SMJ100A
367
Operating Manual 1403.7458.32 ─ 14
Remote command:
[:SOURce<hw>]:BB:DM:CLOCk:MULTiplier
Measured External Clock
Displays the measured frequency of the external clock signal. This enables the user to
permanently monitor the frequency of the externally introduced clock.
This information is displayed only if the external clock source has been selected.
Remote command:
Global Trigger/Clock/Input Settings
Calls the "Global Trigger/Clock/Input Settings" menu. This menu is used among other
things for setting the trigger threshold, the input impedance and the polarity of the clock
and trigger inputs.
The parameters in this menu affect all digital modulations and standards, and are
described in the
Chapter 5.2.3.10, "Global Trigger/Clock/External Input Settings"
Remote command:
n.a.
5.8.3 Arbitrary Waveform Generator ARB
The Arbitrary Waveform Generator (ARB) is an I/Q modulation source forming an inte-
gral part of the R&S
SMJ. It can be used to output any externally calculated modulation
signals or internally generated multi carrier signals or waveform files.
The Arbitrary Waveform Generator is incorporated in the Baseband Generator
(R&S
SMJ-B9/B10/B11) option. The Baseband Main Module (R&S
SMJ-B13) option is
also required. The three Baseband Generator options feature different ARB memory
sizes (see data sheet). Apart from the memory size, however, the three options offer
the same functionality, either one can be installed.
5.8.3.1
Introduction
This chapter provides background information on the ARB functionality and the impact
of the provided settings.
For description on the multi carrier, multi segment and multi carrier continuous wave
functions, refer to:
●
Chapter 5.8.3.4, "Multi Carrier Waveforms"
●
Chapter 5.8.3.3, "Multi Segment Waveforms"
●
Chapter 5.8.4, "Multi Carrier Continuous Wave"
Waveform Files Sources
The R&S
SMJ can replay any waveform file stored on a network drive but also gener-
ate and replay simple waveforms for test purposes.
Baseband Signal - Baseband Block
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...