frames, for example in the course of base station tests, and yet the signal can still
be generated synchronously. In each frame the base station generates a trigger
which would cause a signal generation restart every time but for the suppression.
Example:
Entering 1000 samples means that after a trigger event, any subsequent trigger signal
is ignored for the space of 1000 samples.
5.8.1.8
Marker Output Signals
The R&S
SMJ generates user-definable marker output signals which can be used to
synchronize external instruments. By this means a slot clock or frame clock can be set,
for instance, or the start of a particular modulation symbol can be marked.
Four marker outputs are available. The outputs for markers 1 to 3 are defined, but
marker 4 can be placed on one of the USER outputs according to choice.
Marker
Connector
Location
Markers 1/2
MARKER 1 / 2 BNC connectors
front panel
Marker 3
MARKER 3 pin of the AUX I/O connector
rear panel
Marker 4
USER1 BNC connector or USER2/3/4
pins of the AUX I/O connector
rear panel
Marker Modes
Marker Mode Restart
The generated marker signal is a single "On" pulse. The rising edge of this pulse is
generated at the signal generation start as well as at each subsequent signal restart
time. This marker can be used to monitor the effects of the selected trigger, e.g. trigger
causing restarts of the signal generation.
Marker Mode Pulse
Periodic marker with consecutive On and Off periods of equal length. The first On
period starts at the beginning of the first generated sample/symbol. The marker fre-
quency is defined by a "Divider". The frequency is derived as follows:
<Frequency> = "Symbol/Sample Rate" / "Divider", respectively
<Frequency> = "Sampling Frequency" / "Divider".
Baseband Signal - Baseband Block
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...