Instrument Function
R&S
®
SMJ100A
260
Operating Manual 1403.7458.32 ─ 14
Note:
It is recommended to switch off the display update for optimum sweep performance
especially with short dwell times (see
Chapter 5.2.3.7, "Display Update"
Remote command:
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:DWELl
Global Trigger/Clock Settings
Calls the "Global Trigger/Clock Settings" dialog. This dialog is used among other things
for setting the polarity of the active slope of an applied instrument trigger.
See
Chapter 5.2.3.10, "Global Trigger/Clock/External Input Settings"
User Marker / AUX I/O
Calls the menu for configuring the AUX I/O and user interfaces (see
"User Marker / AUX I/O Settings"
The pin assignment of the AUX I/O interface is shown in the "User Marker / AUX I/O
Configuration" menu; the assignment of the USER pins 2...4 and USER 1 BNC con-
nector can be changed.
5.5.14 RF Sweep and List Mode
5.5.14.1
Overview
The R&S
SMJ offers three different sweep types (frequency sweep, level sweep and
LF sweep) to be activated alternatively. Each type has 6 modes which differ with
respect to the sweep cycle mode (continuous, individual and step-by-step) and trigger-
ing mode (automatic, internal and external).
●
Sweeps and list mode can not be activated simultaneously, they deactivate each
other.
●
If you want to remain at a specific frequency or level value during a sweep, enter
the value directly in the status bar. The sweep stops immediately.
●
A phase continuous frequency sweep can be generated after activating the phase
continuous mode, see
"Setting a Phase Continuous Frequency Sweep"
Setting a sweep
A sweep is set in five basic steps which are shown below taking a frequency sweep as
an example.
1. Set the sweep range ("Start Freq" and "Stop Freq" or "Center Freq" and "Span").
2. Select linear or logarithmic sweep spacing ("Spacing").
3. Set the step width ("Step Lin/Log") and dwell time ("Dwell Time").
4. Activate the sweep ("Mode" to Auto, Single, Step or Extern Single, Extern Step).
RF Signal and Analog Modulations - RF Block
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...