Preparing for Use
R&S
®
SMCV100B
23
Getting Started 1432.7046.02 ─ 02
4.12
Connecting to Dig. IQ HS x
The "Dig. IQ HS x" connector comprises a QSFP+ (Quad Small Form-factor Plug-
gable) socket, that has two components: a QSFP+ cage and a QSFP+ connector.
The QSFP+ cable is equipped with the QSFP+ plug.
1 = QSFP+ plug
2 = QSFP+ cage
3 = QSFP+ connector
The connector is located on the
To connect to Dig. IQ HS x interface
1. For connection, use the QSFP+ cable R&S
DIGIQ-HS.
"Cable selection and electromagnetic interference (EMI)"
2. Hold the QSFP+ plug of the cable by its panes.
3. Turn the QSFP+ cable, so that the release tab shows upwards.
4. Insert and push the QSFP+ plug into the QSFP+ cage.
To disconnect from Dig. IQ HS x interface
1.
NOTICE!
If you pull the cable, you can damage the cable and the "Dig. IQ HS
x" connector.
Pull the release tab.
2. Pull the QSFP+ plug out of the QSFP+ cage.
Connecting to Dig. IQ HS x