RF Signal Configuration
R&S
®
SMA100B
72
User Manual 1178.3834.02 ─ 03
Readjust
Recalculates and adjusts the internal switch positions of the RF chain according to the
current level.
Remote command:
User Variation
Defines and activates a user-defined step width for varying the RF frequency or RF
level with the rotary knob.
If disabled, the step width varies in steps of one unit at the cursor position.
Variation Active ← User Variation
Activates the set user-defined step width.
Remote command:
[:SOURce<hw>]:FREQuency:STEP:MODE
Variation Step ← User Variation
Sets the user-defined step width.
Remote command:
[:SOURce<hw>]:FREQuency:STEP[:INCRement]
[:SOURce<hw>]:POWer:STEP[:INCRement]
3.5 RF Phase Settings
The phase in sinusoidal signals defines the initial angle at its origin.
Access:
1. Select "Frequency" > "Phase".
In the "RF Phase" tab, you can determine the delta phase value and reset the
phase to this reference.
2. Observe the information on the home screen, "Frequency" tile.
The "Frequency" tile indicates used reference frequency, current frequency offset
and multiplier values, and the delta phase value.
RF Phase Settings