Risk of RF connector and cable damage
If you tighten the connectors too strongly, you can damage the cables and connectors.
If you do not tighten the connectors enough, the measurement results can be inaccu-
rate.
Always use an appropriate torque wrench suitable for this type of connector and apply
the torque specified in the application note
The application notes are available on the Internet and provide additional information
on care and handling of RF connectors.
Rohde
&
Schwarz offers appropriate torque wrenches for various connectors. For
ordering information, see the R&S
SMA100B data sheet or product brochure.
2.2.1.12
Pulse Signal Connectors
Pulse Sync
Output signal for synchronizing the pulse generator signal. The synchronization signal
is generated at the beginning of each pulse. For double-pulse generation, the synchro-
nization signal is generated at the beginning of the first pulse.
PULSE VIDEO
Output of the internal pulse generator signal or the external pulse signal fed in via the
PULSE EXT connector (video signal).
See
Chapter 4.4.1, "Pulse Modulation"
PULSE EXT
Input for an external pulse modulation signal or an external trigger/gate signal for the
pulse generator.
2.2.1.13
LF Modulation Connectors
LF
Output for internal LF generator signal.
See also data sheet and user manual, section "Analog Modulation".
EXT 1/2
Inputs for external analog modulation signal, and an external detector voltage.
See
Chapter 4, "Analog Modulations"