Remote Control Commands
R&S
®
SGT100A
485
User Manual 1176.8674.02 ─ 07
:SYSTem:KLOCk
<State>
K
eyboard
LOC
k disables the front panel keys of the instrument.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
SYST:KLOC ON
locks the front panel and external controls. To enable the con-
trols, set
SYST:KLOC OFF
.
:SYSTem:LANGuage
<Language>
Sets the command set to be used.
The instrument can also be remote controlled via the command set of several other
generators.
Note:
While working in a emulation mode, the instrument's specific command set is
disabled, i.e. the SCPI command
SYSTem:LANGuage
will be discarded.
The return to the SCPI command set of the R&S
SGT can only be performed by using
the appropriate command of the selected command set.
Parameters:
<Language>
string
Example:
SYSTem:LANGuage "SCPI"
sets the SCPI command set.
Manual operation:
See
:SYSTem:STARtup:COMPlete?
Queries if the startup of the instrument is completed.
Return values:
<Complete>
0 | 1 | OFF | ON
*RST:
0
Example:
SYST:STAR:COMP?
// 1
// the startup of the instrument is completed
Usage:
Query only
:SYSTem:PROTect<ch>[:STATe]
<State>[, <Key>]
Activates/deactivates the specified protection level.
Parameters:
<State>
select
*RST:
1
SYSTem Subsystem