Remote Control Commands
R&S
®
SGT100A
482
User Manual 1176.8674.02 ─ 07
Parameters:
<Gateway>
string
Range:
0.0.0.0 to ff.ff.ff.ff
Example:
SYSTem:COMMunicate:NETWork:IPADdress:GATeway
'1.2.3.4'
sets the IP address of the default gateway.
Manual operation:
See
:SYSTem:COMMunicate:NETWork[:IPADdress]:SUBNet:MASK
<Mask>
Sets the subnet mask.
Parameters:
<Mask>
string
Example:
SYSTem:COMMunicate:NETWork:IPADdress:SUBNet:
MASK '255.255.0.0'
determines the subnet mask.
Manual operation:
See
:SYSTem:COMMunicate:NETWork:RESource?
Queries the VISA resource string, used for remote control of the instrument with
VXI-11 protocol.
Return values:
<Resource>
string
Example:
SYSTem:COMMunicate:NETWork:RESource?
Response:
"TCPIP::192.1.2.3::INSTR"
Usage:
Query only
Manual operation:
See
:SYSTem:COMMunicate:SERial:BAUD
<Baud>
Sets the baudrate for the serial remote control interface.
Parameters:
<Baud>
2400 | 4800 | 9600 | 19200 | 38400 | 57600 | 115200
*RST:
115200
Example:
SYSTem:COMMunicate:SERial:BAUD 115200
determines 115200 baudrate.
:SYSTem:COMMunicate:SERial:PARity
<Parity>
Sets the parity for the serial remote control interface.
SYSTem Subsystem