Remote Control Commands
R&S
®
SGT100A
353
User Manual 1176.8674.02 ─ 07
[:SOURce<hw>]:BB:ARBitrary:CLOCk
<Clock>
The command sets the clock rate in samples. Loading a waveform waveform
(
ARB:WAV:SEL <name>
) sets the clock rate that is defined in the waveform tag 'clock'.
The command subsequently changes the clock rate; see data sheet for value range.
In the case of an external clock source (selection
ARB:CLOCk:SOURce EXTernal
)
the clock for the external source must be entered with this command.
Parameters:
<Clock>
float
Increment: 0.001
Example:
BB:ARB:CLOC:SOUR INT
selects the internal clock source for generating waveforms.
BB:ARB:CLOC 0.5 MHz
sets the clock rate to 0.5 MHz.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:CLOCk:MODE
<Mode>
The command enters the type of externally supplied clock (
:BB:ARB:CLOCk:SOURce
EXTernal
). When MSAMple is used, a multiple of the sample clock is supplied via the
CLOCK connector and the sample clock is derived internally from this. The multiplier is
entered with the command
:BB:ARB:CLOCk:MULTiplier
.
Parameters:
<Mode>
SAMPle | MSAMple
*RST:
SAMPle
Example:
BB:ARB:CLOC:MODE SAMP
selects clock type "Sample", i.e. the supplied clock is a sample
clock.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:CLOCk:MULTiplier
<Multiplier>
The command specifies the multiplier for clock type "Multiple Samples"
(
:BB:ARB:CLOCk:MODE MSAM
) in the case of an external clock source.
Parameters:
<Multiplier>
integer
Range:
1 to 64
*RST:
4