Remote Control Commands
R&S
®
SGT100A
349
User Manual 1176.8674.02 ─ 07
Example:
SOURce:AWGN:MODE ADD
selects Additive Noise mode.
SOURce:AWGN:DISP:MODE RF
sets the display mode to RF
SOURce:AWGN:POWer:RMODe NOISe
selects Reference Mode Noise. The noise power is kept con-
stant when the C/N value or Eb/N0 value is changed. The carrier
power is adjusted.
Manual operation:
See
[:SOURce<hw>]:AWGN:POWer:SUM?
In Additive Noise and CW Interferer (
SOUR:AWGN:MODE ADD|CW
) modes, queries the
overall power of the noise signal plus useful signal.
Return values:
<Sum>
float
Range:
-145 to 20
Increment: 0.01
*RST:
0
Example:
SOURce:AWGN:POWer:SUM?
queries the overall power of the noise signal plus useful signal.
Usage:
Query only
Manual operation:
See
"C Noise Power / Noise Power (System
Bandwidth) / C Interferer Power / Interferer
Power "
[:SOURce<hw>]:AWGN:POWer:SUM:PEP?
In Additive Noise and CW Interferer (
SOUR:AWGN:MODE ADD|CW
) modes, queries the
peak envelope power of the overall signal comprised of noise signal plus useful signal.
Return values:
<Pep>
float
Range:
-145 to 20
Increment: 0.01
*RST:
0
Example:
SOURce:AWGN:POWer:SUM:PEP?
queries the peak envelope power of the overall signal.
Usage:
Query only
Manual operation:
See
"C Noise PEP / Noise PEP (Total Band-
width) / C Interferer PEP / Interferer PEP"
SOURce Subsystem