Instrument setup
R&S
®
RTO6
98
User Manual 1801.6687.02 ─ 05
2. Select the "Touchscreen" tab.
3. Tap "Calibrate".
A blinking cross appears in the lower left corner of the screen.
4. Touch and hold the cross until "OK" is shown.
5. Repeat this action for the crosses in the other corners.
6. Tap the Rohde
&
Schwarz logo to display the instrument's user interface.
4.6
Preset setup
A user-defined preset contains the complete instrument setup including display set-
tings, except for transparency and intensity. You can save the current configuration to a
preset file, and load a previously saved preset file. You can then specify that these set-
tings are to be applied with the [Preset] key.
4.6.1
User-defined preset
Access: "Menu" > "Settings" > "Preset" > "Settings" tab
Preset
Performs a user-defined preset of the instrument.
Saveset to be loaded on preset
The file name with extension
.dfl
to load or to save the settings to.
Remote command:
Preset setup