![R&S RTO6 Скачать руководство пользователя страница 888](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646888.webp)
Protocol analysis
R&S
®
RTO6
888
User Manual 1801.6687.02 ─ 05
Remote command:
Data Type
Searches for a specific data type. The data type setup consists of the condition and
one or two data patterns.
"Condition"
Defines the operator to set a specific data type ("Equal" or "Not
equal") or a data type range.
"Data Min/Data"
Defines the bit pattern of the data pattern.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
"Data Max"
The second data pattern is required to specify a range with conditions
"In range" and "Out of range".
Remote command:
SEARch:TRIGger:DPHY:DTYPe:CONDition
Word Count/ Data
Searches for a specific word count /data. The word count setup consists of the condi-
tion and one or two data patterns.
"Condition"
Defines the operator to set a specific word count ("Equal" or "Not
equal") or a word count range.
"Data Min/Data"
Defines the bit pattern of the word pattern.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
"Data Max"
The second word pattern is required to specify a range with condi-
tions "In range" and "Out of range".
Remote command:
SEARch:TRIGger:DPHY:WCOunt:CONDition
SEARch:TRIGger:DPHY:WCOunt:MAX
SEARch:TRIGger:DPHY:WCOunt:MIN
Data Value
Searches for a specific data value. The data value setup consists of the condition and
one or two data value patterns.
"Condition"
Defines the operator to set a specific data value ("Equal" or "Not
equal") or a data value range.
D-PHY (option R&S