![R&S RTO6 Скачать руководство пользователя страница 471](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646471.webp)
Data and file management
R&S
®
RTO6
471
User Manual 1801.6687.02 ─ 05
Measurement
Selects the measurement to be exported.
Remote command:
Source, Measurement scale
Displays the measurement settings source and scale.
Export type
You can export the result data of the long-term measurement, the measurement histo-
gram or the track.
To export the data, the required type must be enabled in the ""Menu" > "Measure" >
Plot" tab: "Long term > Enable", "Histogram > Enable" or "Track > Enable".
Remote command:
Incidence
Sets the mode of exported histogram data: relative or absolute count of values. If rela-
tive values are exported, the sum of all values is 1, and the count of each value is set
in relation to the sum.
Remote command:
Save to file
Enter the filename to save the measurement data to. Double-tap the filename to open
the file selection dialog box.
By default, the filename has the prefix "Measurement_".
You can define a pattern for automatic naming in the "Autonaming" tab, see
The default directory is:
C:\Users\Public\Documents\Rohde-Schwarz\RTO6\Measurements
"Save"
Saves the measurement data in the selected file.
"Save As..."
Opens the file selection dialog box and saves the measurement data
to the selected file. See also
Chapter 12.5, "File selection dialog"
".bin/.xml/.csv"
Selects the file format.
Remote command:
12.2.6
Eyes
Access: "Menu" key >"Save/Recall" key > "Save" tab > "Eyes"
Save and recall waveform data and results