Remote control commands
R&S
®
RTO6
2509
User Manual 1801.6687.02 ─ 05
ADVJitter<m>:DCOMposition:ENP
<State>
Enables the display of the periodic jitter components table.
Suffix:
<m>
.
1..2
Parameters:
<State>
ON | OFF
*RST:
ON
ADVJitter<m>:DCOMposition:NHORizontal
<MaxNumber>
Sets the maximum number of periodic components in the horizontal direction that are
considered for the measurement. If the actual number of periodic components is higher
than the set value, they are considered for the random jitter component.
Suffix:
<m>
.
1..2
Parameters:
<MaxNumber>
Range:
0 to 10
Increment: 1
*RST:
5
ADVJitter<m>:DCOMposition:NVERtical
<MaxNumber>
Sets the maximum number of periodic components in the vertical direction that are
considered for the measurement. If the actual number of periodic components is higher
than the set value, they are considered for the random jitter component.
Suffix:
<m>
.
1..2
Parameters:
<MaxNumber>
Range:
0 to 10
Increment: 1
*RST:
5
ADVJitter<m>:DCOMposition:ENOBu
<State>
Suffix:
<m>
.
1..2
Parameters:
<State>
ON | OFF
*RST:
OFF
Jitter analysis and clock data recovery