Remote control commands
R&S
®
RTO6
2473
User Manual 1801.6687.02 ─ 05
IQ:STATe
<IQMode>
Activates the I/Q mode of the instrument.
Parameters:
<IQMode>
ON | OFF
ON: I/Q mode
OFF: normal oscilloscope mode
*RST:
OFF
Usage:
Asynchronous command
IQ:INSRate
<InputSampleRate>
Shows the sample rate of capturing the input data in I/Q mode: 10
Gsample/s.
Return values:
<InputSampleRate> SR10G
Usage:
Asynchronous command
CHANnel<m>:IQ:INPType
<IQInputType>
Sets the type of the input signal.
Suffix:
<m>
.
1..4
Parameters:
<IQInputType>
REAL | COMPlex
REAL
Real RF signal. One real RF signal requires one input channel,
thus up to four real signals can be recorded in parallel.
COMPlex
Complex I/Q signal in baseband or low IF range. One complex
input signal requires two input channels. The In-Phase compo-
nent must be connected to channel 1 or 3, and the Quadrature
component must be connected to channel 2 or 4. Thus up to two
complex input signals can be recorded in parallel.
*RST:
REAL
Usage:
Asynchronous command
CHANnel<m>:IQ:INPMode
<IQInputMode>
Selects the frequency band of a complex input signal.
Suffix:
<m>
.
1..4
Selects the input channel.
I/Q software interface (option R&S