Remote control commands
R&S
®
RTO6
2036
User Manual 1801.6687.02 ─ 05
Query parameters:
<SearchName>
String parameter
Return values:
<FrameCount>
Range:
0 to 100000
Increment: 1
*RST:
0
Usage:
Query only
SEARch:RESult:ETHernet:WORD<m>:STATe?
<SearchName>
Returns the frame state of the selected word within the search result.
Suffix:
<m>
.
*
Query parameters:
<SearchName>
String parameter
Return values:
<State>
OK | ERR_PREAMBLE | ERR_LENGTH
OK
No error detected
ERR_PREAMBLE
Error in the preamble of the selected word
ERR_SFD
Error in the start frame delimiter (SFD). The value of a correct
SFD byte is 171. The SFD is transmitted LSB first.
ERR_LENGTH
Error in the number of bits in the selected word
*RST:
OK
Usage:
Query only
SEARch:RESult:ETHernet:WORD<m>:STARt?
<SearchName>
Returns the frame start time of the selected word within the search result.
Suffix:
<m>
.
*
Query parameters:
<SearchName>
String parameter
Return values:
<Start>
Range:
-100E+24 to 100E+24
Increment: 100E-12
*RST:
0
Default unit: s
Usage:
Query only
Protocols