Remote control commands
R&S
®
RTO6
1900
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Threshold>
Range:
-12 to 12
Increment: 0.1
*RST:
1.65
Default unit: V
BUS<m>:SENT:TECHnology
<Technology>
Selects the threshold voltage.
Suffix:
<m>
.
1..4
Selects the serial data bus.
Parameters:
<Technology>
V25 | MANual
V25
The threshold value is 2.5 V, according to CMOS technology.
MANual
Sets the threshold to the value set with
*RST:
V25
BUS<m>:SENT:CLKPeriod
<ClockPeriod>
Sets the nominal clock period (clock tick).
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<ClockPeriod>
Range:
1E-6 to 100E-6
Increment: 1E-6
*RST:
6E-6
Default unit: s
BUS<m>:SENT:CLKTolerance
<ClockTolerance>
Sets a tolerated deviation of the clock signal.
Suffix:
<m>
.
1..4
Selects the serial bus.