![R&S RTO6 Скачать руководство пользователя страница 1895](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461895.webp)
Remote control commands
R&S
®
RTO6
1895
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
*
Query parameters:
<SearchName>
Return values:
<HeadCksumState> OK | STERror | SPERror | PRERror | UVAL | INSufficient
STERror: StarT ERror, incorrect start bit
SPERror: StoP ERror, incorrect stop bit
PRERror:PaRity ERror, incorrect parity bit.
UVAL: unexpected value
*RST:
OK
Usage:
Query only
SEARch:RESult:FLXRay:FRAMe<m>:CSValue?
<SearchName>
Returns the value of the header checksum.
Suffix:
<m>
.
*
Query parameters:
<SearchName>
Return values:
<HeadCksumValue> Range:
0 to 2047
Increment: 1
*RST:
0
Usage:
Query only
SEARch:RESult:FLXRay:FRAMe<m>:CYCount?
<SearchName>
Suffix:
<m>
.
*
Query parameters:
<SearchName>
Return values:
<CycleCount>
Range:
0 to 65535
Increment: 1
*RST:
0
Usage:
Query only
SEARch:RESult:FLXRay:FRAMe<m>:DATA?
<SearchName>
Suffix:
<m>
.
*
Protocols