![R&S RTO6 Скачать руководство пользователя страница 1487](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461487.webp)
Remote control commands
R&S
®
RTO6
1487
User Manual 1801.6687.02 ─ 05
Parameters:
<Identifier>
ON | OFF
*RST:
OFF
Example:
:MEASurement:ARNames 1
:MEASurement:ARES?
Low: -1.185770750988e-001,Amplitude:
2.371541501976e-001,Max: 1.343873517787e-001
Usage:
Asynchronous command
MEASurement<m>:RESult[:ACTual]?
[<MeasType>]
MEASurement<m>:RESult:AVG?
[<MeasType>]
MEASurement<m>:RESult:EVTCount?
[<MeasType>]
MEASurement<m>:RESult:NPEak?
[<MeasType>]
MEASurement<m>:RESult:PPEak?
[<MeasType>]
MEASurement<m>:RESult:RELiability?
[<MeasType>]
MEASurement<m>:RESult:RMS?
[<MeasType>]
MEASurement<m>:RESult:WFMCount?
[<MeasType>]
MEASurement<m>:RESult:STDDev?
[<MeasType>]
Return the statistic results of the specified measurement. If no parameter is specified,
the result of the main measurement is returned. The main measurement is defined
using
.
●
[:ACTual]: current measurement result
●
AVG: average of the long-term measurement results
●
EVTCount: number of measurement results in the long-term measurement
●
NPEak: negative peak value of the long-term measurement results
●
PPEak: positive peak value of the long-term measurement results
●
RELiability: reliability of the measurement result
●
RMS: RMS value of the long-term measurement results
●
STDDev: standard deviation of the long-term measurement results
For a detailed description of the results, see
"Selection of the measurement group:
Automatic measurements