Remote control commands
R&S
®
RTO6
1433
User Manual 1801.6687.02 ─ 05
Parameters:
<RunExecutable>
NOACtion | TRIGger
*RST:
NOACtion
Usage:
Asynchronous command
TRIGger<m>:OUT:ACTion
<StateAction>
Enables/disbales the trigger out signal that is provided to the [Trigger Out] connector
on the rear panel when a trigger occurs.
Suffix:
<m>
.
1..3
Parameters:
<StateAction>
NOACtion | TRIGger
*RST:
NOACtion
Usage:
Asynchronous command
24.9.21
Zone trigger
TRIGger<m>:ZONE:EXPRession[:DEFine]
<LogiExpr>
Defines the zone trigger.
Suffix:
<m>
.
1..3
The suffix is irrelevant.
Parameters:
<LogiExpr>
String with the logical expression
Example:
Prerequisite: mask tests MT1 and MT2 are defined, and also
Zone1.
TRIGger:ZONE:EXPRession 'MT1 and MT2 and Zone1'
TRIGger:ZONE:EXPRession?
<-- MT1 and MT2 and Zone1
Usage:
Asynchronous command
Firmware/software:
FW 3.20
SWTRigger:HISTory
<ApplyZnTrigHistory>
Applies the zone trigger condition to the acquisitions in the history memory during his-
tory replay.
"Apply zone trigger to history"
Parameters:
<ApplyZnTrigHistory> ON | OFF
*RST:
OFF