Remote control commands
R&S
®
RTO6
1415
User Manual 1801.6687.02 ─ 05
Usage:
Asynchronous command
Standard
Minimum value
Maximum value
PAL
1
625
PAL-M
1
525
NTSC
1
263 in odd field
262 in even field
SECAM
1
625
480p/60 (
P480L60HZ
)
1
525
720p/30 (
P720L30HZ
)
720p/50 (
P720L50HZ
)
720p/60 (
P720L60HZ
)
1
750
1080i/50 (
I1080L50HZ
)
1080i/60 (
I1080L60HZ
)
1080p/24 (
P1080L24HZ
)
1080p/24sF (
P1080L24HZSF
)
1080p/25 (
P1080L25HZ
)
1080p/30 (
P1080L30HZ
)
1080p/50 (
P1080L50HZ
)
1080p/60 (
P1080L60HZ
)
1
1125
TRIGger<m>:TV:LFIeld
<LineField>
The commands is only relevant for NTSC signals and sets the field in which the line
number is counted.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<LineField>
FIELD1 | FIELD2
FIELD1 = odd field
FIELD2 = even field
*RST:
FIELD1
Usage:
Asynchronous command
TRIGger<m>:TV:CUSTom:SCANmode
<ScanMode>
Sets the scanning system. Only relevant if
CUSTom
.
See also:
Trigger