Power analysis (option R&S
RTO6-K31)
R&S
®
RTO6
1166
User Manual 1801.6687.02 ─ 05
Table 19-3: Results of the mask test
Result
Description
Acq. completed
Number of tested acquisitions
Acq. remaining
Remaining acquisitions until "Average count / Nx
Single count" is reached
State
Shows if the test has been completed. The state is
set to "Finished" when "Nx Single count" acquisi-
tions are tested and the number of "Acq. remaining"
is 0. As long as the number of tested acquisitions is
less the "Nx Single count" number, the state is "Run-
ning".
If you run the acquisition with [Run Stop], or the
number of played history acquisitions exceeds "Nx
Single count", the mask testing is performed accord-
ing to fail criteria settings independently of the test
state. The testing is not stopped when the state is
set to "Finished".
Sample hits
Number of samples that hit the mask
Acquisition hits
Number of acquisitions that contained at least one
sample hit
Fail rate
Ratio of acquisition hits to the number of tested
acquisitions
Test result
A test has failed if the number of sample hits or
acquisition hits exceeds the limit of "Violation toler-
ance" hits
The following remote commands are used for handling the measurement results:
●
●
19.5.1.2
Configuring safe operating area
For details of the configuration settings, see
Chapter 19.5.1.3, "Safe operating area
1. Select "Analysis" > "Power".
2. Under "Power Path", select "Safe Operating Area".
3. Connect the differential voltage probe and the current probe to the oscilloscope.
4. Deskew the probes as described in
Chapter 19.1.1.1, "Auto deskew"
5. Connect the probes to the DUT as shown in the "Channels" tab:
Power path