Remote control commands
R&S
®
RTA4000
706
User Manual 1335.7898.02 ─ 08
TRIGger:A:ARINc:LABel:MAXimum
<LabelMaximum>
Specifies the maximum value of the label if
TRIGger:A:ARINc:LABel:CONDition
is set to
INRange
or
OORange
.
Parameters:
<LabelMaximum>
01X-string
TRIGger:A:ARINc:LABel:MINimum
<LabelMinimum>
Sets the minimum value of the label to be triggerd on.
Parameters:
<LabelMinimum>
01X-string
TRIGger:A:ARINc:SDI
<SDIvalue>
Sets the source/destination identifier (SDI) bits to be triggered on.
Parameters:
<SDIvalue>
ANY | S00 | S01 | S10 | S11
TRIGger:A:ARINc:SSM
<SSMvalue>
Sets the sign/status matrix (SSM) bits to be triggered on.
Parameters:
<SSMvalue>
ANY | S00 | S01 | S10 | S11
TRIGger:A:ARINc:TTIMe:CONDition
<Compare>
Define the operators for comparison of the decoded transmission time condition with
the specified data.
Parameters:
<Compare>
GTHan | LTHan | WITHin | OUTSide
*RST:
GTH
TRIGger:A:ARINc:TTIMe:MAXimum
<TransmissionTimeMax>
Sets the maximum value of the transmission time.
Parameters:
<TransmissionTimeMax>
TRIGger:A:ARINc:TTIMe:MINimum
<TransmissionTimeMin>
Sets the minimum value of the transmission time.
Serial bus analysis