Remote control commands
R&S
®
RTA4000
653
User Manual 1335.7898.02 ─ 08
Parameters:
<IdType>
B11 | B29
*RST:
B11
SEARch:PROTocol:CAN:ICONdition
<IdCondition>
Sets the comparison condition for the identifier: If the pattern contains at least one X
(don't care), you can trigger on values equal or not equal to the specified value. If the
pattern contains only 0 and 1, you can also trigger on a range greater than or lower
than the specified value.
is set to
IDENtifier
,
IDData
, or
IDERror
.
Parameters:
<IdCondition>
EQUal | NEQual | GTHan | LTHan
*RST:
EQU
SEARch:PROTocol:CAN:IDENtifier
<Identifier>
Defines the identifier pattern. The pattern length is defined with
.
is set to
IDENtifier
,
IDData
, or
IDERror
.
Parameters:
<Identifier>
String containing binary pattern with max. 29 bit. Characters 0,
1, and X are allowed.
SEARch:PROTocol:CAN:DLENgth
<DataLength>
Defines the length of the data pattern - the number of bytes in the pattern.
is set to
IDData
.
Parameters:
<DataLength>
Range:
0 to 8
Increment: 1
*RST:
1
Default unit: Byte
SEARch:PROTocol:CAN:DCONdition
<DataCondition>
Sets the comparison condition for data: If the pattern contains at least one X (don't
care), you can trigger on values equal or not equal to the specified value. If the pattern
contains only 0 and 1, you can also trigger on a range greater than or lower than the
specified value.
is set to
IDData
.
Serial bus analysis