![R&S RTA4000 Скачать руководство пользователя страница 631](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987631.webp)
Remote control commands
R&S
®
RTA4000
631
User Manual 1335.7898.02 ─ 08
Suffix:
<b>
.
1..4
Parameters:
<Polarity>
POSitive | NEGative
POSitive = high active
NEGative = low active
*RST:
POS
BUS<b>:UART:SSIZe
<SymbolSize>
Sets the number of data bits in a message.
Suffix:
<b>
.
1..4
Parameters:
<SymbolSize>
Range:
5 to 9
Increment: 1
*RST:
8
Default unit: Bit
BUS<b>:UART:PARity
<Parity>
Defines the optional parity bit that is used for error detection.
Suffix:
<b>
.
1..2
Selects the bus.
Note: SPI and UART protocols occupy two bus lines.
Parameters:
<Parity>
ODD | EVEN | NONE
*RST:
NONE
BUS<b>:UART:SBIT
<StopBitNumber>
Sets the stop bits.
Suffix:
<b>
.
1..2
Selects the bus.
Note: SPI and UART protocols occupy two bus lines.
Parameters:
<StopBitNumber>
B1 | B1_5 | B2
1 stop bit, 1.5 stop bits or 2 stop bits are possible.
*RST:
B1
Serial bus analysis