Remote control commands
R&S
®
RTA4000
617
User Manual 1335.7898.02 ─ 08
<n>
*
Selects the frame.
Return values:
<DataMiso>
List of decimal values of data bytes
Example:
BUS:SPI:FRAM3:DATA:MISO?
-> 94,177,171,60,242,219,100,0
Usage:
Query only
BUS<b>:SPI:FRAME<n>:WCOunt?
Returns the number of words in the specified frame.
Suffix:
<b>
.
1..4
<n>
*
Selects the frame.
Return values:
<WordCount>
Number of words
Usage:
Query only
BUS<b>:SPI:FRAME<n>:WORD<o>:STARt?
Returns the start time of the specified data word.
Suffix:
<b>
.
1..4
<n>
*
Selects the frame.
<o>
*
Selects the word number.
Return values:
<StartTime>
Range:
depends on sample rate, record length, and time
base
Increment: depends on the time base
Default unit: s
Usage:
Query only
BUS<b>:SPI:FRAME<n>:WORD<o>:STOP?
Returns the end time of the specified data word.
Suffix:
<b>
.
1..4
Serial bus analysis