![R&S RTA4000 Скачать руководство пользователя страница 609](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987609.webp)
Remote control commands
R&S
®
RTA4000
609
User Manual 1335.7898.02 ─ 08
Parameters:
<MisoSource>
CH1 | CH2 | CH3 | CH4 | NONE | D0..D15
*RST:
NONE
BUS<b>:SPI:DATA:POLarity
<Polarity>
Selects whether transmitted data is high active (high = 1) or low active (low = 1) on the
data line.
Suffix:
<b>
.
1..4
Parameters:
<Polarity>
POSitive | NEGative
POSitive = high active
NEGative = low active
*RST:
POSitive
BUS<b>:SPI:MOSI:POLarity
<MosiPolarity>
Selects if transmitted data is high active (high = 1) or low active (low = 1) on the MOSI/
MISO line.
Suffix:
<b>
.
1..4
Parameters:
<MosiPolarity>
ACTLow | ACTHigh
*RST:
ACTH
BUS<b>:SPI:MISO:POLarity
<MisoPolarity>
Selects whether transmitted data is high active (high = 1) or low active (low = 1) on the
MISO line.
Suffix:
<b>
.
1 | 3. Bus 2 or 4 is not available if the MISO source is used on
bus 1 or 3, respectively.
Parameters:
<MisoPolarity>
ACTLow | ACTHigh
*RST:
ACTH
BUS<b>:SPI:BORDer
<BitOrder>
Defines if the data of the messages starts with MSB (most significant bit) or LSB (least
significant bit).
Suffix:
<b>
.
1..4
Serial bus analysis