Remote control commands
R&S
®
RTA4000
519
User Manual 1335.7898.02 ─ 08
Suffix:
<m>
.
1..8
Selects the measurement place.
Parameters:
<SignalSlope>
POSitive | NEGative
Slope of first waveform ("Measure Source").
*RST:
POS
<ReferenceSlope>
POSitive | NEGative
Slope of the second waveform ("Measure Source 2"). The value
is ignored in delay to trigger measurements.
*RST:
POS
MEASurement<m>:DELay:DIRection
<Direction>,<Direction>
Selects the edges for delay measurement and delay to trigger measurement.
See also:
Suffix:
<m>
.
1..8
Selects the measurement place.
Parameters:
<Direction>
NEARest | FRFI | FRLA
For
delay measurement
, it defines the edge of the first wave-
form ("Measure Source").
NEARest: Nearest edge to the trigger point.
FRFI: First edge from the left window edge.
FRLA: First edge from the right window edge, which is the last
edge if counted from the left window edge.
For
delay to trigger measurement
, it defines the measured
edge.
NEARest: Nearest edge to the trigger point.
FRFI: Next edge from the trigger to the right.
FRLA: Next edge from the trigger to the left.
<Direction>
NEARest | FRFI | FRLA
For delay measurement, it defines the edge of the second wave-
form ("Measure Source 2").
NEARest: Nearest edge to the edge determined on "Measure
Source".
FRFI: Next edge to the right, counted from the edge determined
on "Measure Source".
FRLA: Next edge to the left, counted from the edge determined
on "Measure Source"
For delay to trigger measurement, enter any value. This value is
ignored.
Measurements