![R&S RTA4000 Скачать руководство пользователя страница 437](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987437.webp)
Remote control commands
R&S
®
RTA4000
437
User Manual 1335.7898.02 ─ 08
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Range>
Voltage range value
Range:
8e-3 to 80 (without probe attenuation)
*RST:
40e-3
Default unit: V
CHANnel<m>:POSition
<Position>
Sets the vertical position of the waveform in divisions. While the offset sets a voltage,
position is a graphical setting given in divisions.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Position>
Graphical position
Range:
-5 to 5
*RST:
0
Default unit: div
CHANnel<m>:OFFSet
<Offset>
Sets the offset voltage, which is subtracted to correct an offset-affected signal.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Offset>
Offset value
Range:
Depend on vertical scale and probe attenuation.
Increment: Depends on vertical scale and probe attenuation.
*RST:
0
Default unit: V
CHANnel<m>:COUPling
<Coupling>
Selects the connection of the indicated channel signal - coupling and termination.
Waveform setup