Network connections and remote operation
R&S
®
RTA4000
243
User Manual 1335.7898.02 ─ 08
12.2
USB connection
Alternatively to a LAN connection, you can use the USB type B (device USB) connec-
tor at the rear panel to access the instrument via USB.
Connecting the instrument using USB
1. Connect the USB cable to the USB type B connector at the rear panel of the instru-
ment, and to the computer.
2. Open the "Setup" menu.
3. Tap "Interface".
4. Tap "USB" to select USB connection.
5. Tap "Parameter".
6. Select the USB mode.
● USB TMC (Test & Measurement Class)
● USB VCP (Virtual Com Port)
● USB MTP (Media Transfer Protocol)
12.2.1
USB TMC
USB TMC means USB Test & Measurement Class. You can use this interface for
remote control of the instrument using SCPI commands. USB TMC does not need a
driver installation but requires a VISA installation on the controlling computer. VISA is
used to access the instrument, to send remote commands, and to read status informa-
tion.
Rohde
&
Schwarz provides the standardized I/O software library R&S
VISA for com-
munication via TCP/IP (LAN: HiSlip, VXI-11) or USB (USBTMC) interfaces. R&S
VISA
is available for download at the Rohde
&
.
USB connection