Preparing for use
R&S
®
QAR50
20
User Manual 1179.4992.02 ─ 02
To suppress electromagnetic radiation during operation:
●
Use high-quality shielded cables, for example, double-shielded RF and LAN
cables.
●
Always terminate open cable ends.
●
Ensure that connected external devices comply with EMC regulations.
4.5
Connecting the R&S
QAR50
The following overview shows the possible connections for the R&S
QAR50.
R&S QAR50
AC power
Monitor (DisplayPort)
LAN
Keyboard / mouse (USB)
Figure 4-12: Connections for the R&S
QAR50
1. Connect a monitor to the DisplayPort port on the R&S
QAR50.
As a power source, you can use a household AC outlet to supply the monitor with
power (for monitors with a light-current extension cable).
Note:
We recommend using a monitor with a resolution of at least 1920 x 1080.
Lower resolutions can lead to an incorrect display of the measurement data.
2. Connect a keyboard and a mouse to two of the USB ports.
3. Optional: Establish a connection to your LAN via the LAN interface.
Note:
Always connect all devices before connecting the R&S
QAR50 to the mains
supply.
Connecting to power
The R&S
QAR50 contains an AC power supply on the connector board.
For safety information, see
1. Plug the AC power cable into the AC power connector on the connector board.
Only use the AC power cable delivered with the product.
2. Plug the AC power cable into a power outlet with ground contact.
The required ratings are listed next to the AC power connector and in the data
sheet.
Connecting the R&S
QAR50