R&S OSP
Appendix
Interfaces and Connectors
Operating Manual 1505.3896.12 - 14
196
The CAN bus connector is placed in the upper right corner of the R&S OSP rear side.
The connector type is D-Sub, 9-pin, female type.
1
2
4
5
3
9
8
7
6
Figure 9-1: Pin Assignment CAN Bus Connector
Pin
Signal
Type
Remark
1
Not connected
2
CAN-L
bi-directional
maximum 1 Mbit/s
3
GND
4
Not connected
5
Not connected
6
Not connected
7
CAN-H
bi-directional
maximum 1 Mbit/s
8
Not connected
9
Power supply
output
12 V, max. 300 mA
CAN bus connection
When connecting equipment to the CAN bus interface like fiber-optic CAN bus
converters, ensure that both devices support the same transfer rate.
Use the cable R&S OSP-Z106 to connect the R&S OSP device to the fiber-optic CAN
bus converter.
Use the cables R&S OSP-Z101 or R&S OSP-Z102 to connect two R&S OSP devices.
These cables do not connect the 12 V DC line of the CAN bus.
9.1.2.2 LAN REMOTE
8-pin connector RJ-45 used to connect the R&S OSP to a Local Area Network (LAN).
Refer to Remote Operation in a LAN and LAN Interface. The pin assignment of the
RJ-45 connector supports category 5 and 6 UTP/STP (Unshielded/Shielded Twisted
Pair) cables.
This connector is not available on the R&S OSP150.
It is recommended to use double-shielded LAN cables of category 6 (SSTP).