Remote Control Commands
R&S
®
NRX
370
User Manual 1178.5566.02 ─ 07
OUTPut:RECorder<output>:LIMit:LOWer[:VALue]
<value>
Effective if
REC
,
FREC
or
RREC
.
Sets the measurement value that corresponds to 0
V output voltage.
If you enter a value without unit, the unit is defined by
. For further information, see
Suffix:
<output>
.
1 to 2
BNC connectors at the rear; 1
=
Out
1
/
Trig
Out,
2
=
Trig
In
/
Out
2
Parameters:
<value>
The range depends on the measurement.
Manual operation:
OUTPut:RECorder<output>:LIMit:LOWer:POWer
<value>
Effective if
REC
,
FREC
or
RREC
Sets the power measurement value that corresponds to 0
V output voltage.
If you enter a value without unit, the unit is defined by
. For further information, see
Suffix:
<output>
.
1 to 2
BNC connectors at the rear; 1
=
Out
1
/
Trig
Out,
2
=
Trig
In
/
Out
2
Parameters:
<value>
Range:
-180.0 to +210.0
*RST:
-30.0
Default unit: dBm
Manual operation:
OUTPut:RECorder<output>:LIMit:LOWer:RATio:RCOefficient
<value>
Effective if
REC
,
FREC
or
RREC
Sets the reflection coefficient measurement value that corresponds to 0
V output volt-
age.
Suffix:
<output>
.
1 to 2
BNC connectors at the rear; 1
=
Out
1
/
Trig
Out,
2
=
Trig
In
/
Out
2
Configuring the Analog Signal Ouput and the Trigger Input/Output