Remote Control Commands
R&S
®
NRX
253
User Manual 1178.5566.02 ─ 07
Usage:
Query only
Manual operation:
CALCulate<Measurement>:TRACe:MEASurement:POWer:PULSe:TOP?
Effective for pulse analysis measurements.
Queries the pulse top power level detected by the selected
CALCulate<Measurement>:TRACe:MEASurement:ALGorithm
used as a reference (100
%) to determine other parameter values such as the rising or
falling thresholds.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Query only
Manual operation:
CALCulate<Measurement>:TRACe:MEASurement:POWer:REFerence?
Effective for pulse analysis measurements.
Queries the power level at reference level. See
MEASurement:DEFine:DURation:REFerence
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Query only
CALCulate<Measurement>:TRACe:MEASurement:PULSe:DCYCle?
Effective for pulse analysis measurements.
Queries the duty cycle of the measured power.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Query only
Manual operation:
CALCulate<Measurement>:TRACe:MEASurement:PULSe:DURation?
Effective for pulse analysis measurements.
Queries the time between the first positive edge and the subsequent negative edge of
the pulse, where the edges occur at crossings of the mid threshold.
Measurement Settings and Results
www.allice.de
Allice Messtechnik GmbH