Remote Control Commands
R&S
®
NRX
357
User Manual 1178.5566.02 ─ 05
Parameters:
<Domain>
Example:
SYSTem:COMMunicate:NETWork:COMMon:DOMain ABC.DE
Sets the domain of the network to ABC.DE.
Manual operation:
See
SYSTem:COMMunicate:NETWork[:COMMon]:HOSTname
<Hostname>
Sets the individual hostname of the R&S NRX.
Parameters:
<Hostname>
Example:
SYST:COMM:NETW:COMM:HOST 'power_meter_2'
Sets
power_meter_2
as new hostname.
Manual operation:
See
SYSTem:COMMunicate:INET[:SELF]:DNS:ADDRess
<server>
SYSTem:COMMunicate:NETWork[:IPADdress]:DNS
<DNS>
SYSTem:COMMunicate:NETWork[:IPADdress]:MODE
STATic
is set.
Sets the IP address of the network DNS server.
Parameters:
<DNS>
DNS server IPv4 address, consisting of four blocks separated by
dots.
Range:
0 to 255
Example:
SYST:COMM:NETW:DNS "123.456.0.1"
Sets the IP address of the DNS server to 123.456.0.1.
Manual operation:
See
SYSTem:COMMunicate:INET[:SELF]:GATeway:ADDRess
<gateway>
SYSTem:COMMunicate:NETWork[:IPADdress]:GATeway
<Gateway>
SYSTem:COMMunicate:NETWork[:IPADdress]:MODE
STATic
is set.
Sets the IP address of the default gateway.
Parameters:
<Gateway>
The four parameter form the IP address x.y.z.a.
Range:
0 to 255
Example:
SYST:COMM:NETW:GAT "192.168.10.254"
Sets the IP address of the default gateway to 192.168.10.254.
Manual operation:
See