Remote Control Commands
R&S
®
NRX
338
User Manual 1178.5566.02 ─ 05
If you enter a value without unit, the unit is defined by
. For further information, see
Suffix:
<output>
.
1 to 2
BNC connectors at the rear; 1
=
Out
1
/
Trig
Out,
2
=
Trig
In
/
Out
2
Parameters:
<value>
Manual operation:
See
OUTPut:RECorder<output>:LIMit:LOWer:POWer
<value>
Enter the power measurement value that corresponds to the output voltage of 0
V.
Below this limit, the analog output is set to 0
V.
If you enter a value without unit, the unit is defined by
. For further information, see
Suffix:
<output>
.
1 to 2
BNC connectors at the rear; 1
=
Out
1
/
Trig
Out,
2
=
Trig
In
/
Out
2
Parameters:
<value>
Range:
-180.0 to +210.0
*RST:
-30.0
Default unit: dBm
Manual operation:
See
OUTPut:RECorder<output>:LIMit:LOWer:RATio:RCOefficient
<value>
Enter the reflection coefficient measurement value that corresponds to the output volt-
age of 0
V. Below this limit, the analog output is set to 0
V.
Suffix:
<output>
.
1 to 2
BNC connectors at the rear; 1
=
Out
1
/
Trig
Out,
2
=
Trig
In
/
Out
2
Parameters:
<value>
Range:
-1e18 to 1e18
*RST:
0.0
Default unit: -
Manual operation:
See
Configuring the Analog Signal Ouput and the Trigger Output